0102030405
China Suppliers ChallengePore® PES Membrane Filter - High Mycoplasma Retention, Double-Layer Design Factory
Typical Applications
Filtration of cell culture media
Intermediate product and drug substance filtration
Filtration of serum and blood products
Final product filtration
Product Specifications
| Max. Operating Temp | 80°C |
|---|---|
| Max. Operating DP |
0.55 MPa @ 25°C 0.1 MPa @ 80°C |
Biological Safety
Endotoxin: < 0.25 EU/mL
Biocompatibility: Complies with USP <87>, No potential cytotoxicity.
Complies with USP <88> Class VI Biological Reactivity requirements for plastics.
Complies with USP <88> Class VI Biological Reactivity requirements for plastics.
Integrity Test: Bubble Point and Diffusion Flow
Sterilization Methods
ChallengePore® PES Capsule Filters:
Autoclave (121°C, 30 min, 5 cycles) or Gamma Irradiation (25-40 kGy).
Autoclave (121°C, 30 min, 5 cycles) or Gamma Irradiation (25-40 kGy).
ChallengePore® PES Cartridge Filters:
Autoclave (up to 130°C, 30 min, 25 cycles) and Steam-in-Place (SIP).
Autoclave (up to 130°C, 30 min, 25 cycles) and Steam-in-Place (SIP).
Frequently Asked Questions
What are the primary applications of ChallengePore® PES filters?
They are designed for the filtration of cell culture media, intermediate products and drug substances, serum and blood products, and final product filtration.
What are the operating temperature and differential pressure limits?
The maximum operating temperature is 80°C. The maximum operating differential pressure (DP) is 0.55 MPa at 25°C and 0.1 MPa at 80°C.
What biological safety certifications do these filters comply with?
They feature bacterial endotoxin levels < 0.25 EU/mL, satisfy USP <87> with no potential cytotoxicity, and meet USP <88> Class VI Biological Reactivity standards for plastics.
Which integrity testing methods are compatible?
Filter integrity can be verified through both Bubble Point and Diffusion Flow test methods.
How do sterilization protocols differ between capsule and cartridge formats?
Capsule filters support autoclaving (121°C, 30 min, up to 5 cycles) or Gamma Irradiation (25-40 kGy). Cartridge filters support autoclaving (up to 130°C, 30 min, up to 25 cycles) as well as Steam-in-Place (SIP).